FDA Adverse Event Injury Summary report: N

AUTOSOFT XC

MDR report key: 19848506 · Received July 28, 2024

Report

Report Number
3003442380-2024-17492
Event Type
Injury
Date Received
July 28, 2024
Date of Event
June 20, 2024
Report Date
August 6, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
PMA / PMN Number
K032854
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR 1924856 - MDR 3003442380-2024-17492 - DEVICE 1 OF 2.

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2024-17492. CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS - H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN FOR SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE. THE BATCH 6004944 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2 FOR THE CODE SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE COMPLAINT INVESTIGATIONS. PHOTO/SAMPLE WAS NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL FLOW TEST 1 ACCORDING TO WI VERSION 2 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6004944 WAS MANUFACTURED ACCORDING TO THE WI VERSION 110 MANUFACTURED IN THE INSET 9, ON 08/JAN/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 04-AUG-2025 AGAINST MALFUNCTION CODE SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE AND LOT 6004944 AND 2 COMPLAINTS HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT 6004944 AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, 2 COMPLAINTS RECEIVED ON THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT FACED TWO INFUSION SET CANNULA BENT EVENT ON (B)(6) 2024. THE PATIENT WENT TO EMERGENCY ROOM (ER) AND EVENTUALLY GOT HOSPITALIZED DUE TO HIGH BG. THE GLUCOSE LEVEL WAS 394 MG/DL AT THE TIME OF INCIDENT AND THE HIGHEST BLOOD GLUCOSE NOTICED DUE TO THE ISSUE WAS 506 MG/DL LEADING TO DIABETIC KETOACIDOSIS (DKA). THE PATIENT WAS TREATED WITH BOLUS VIA PUMP. UNOMEDICAL DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1587706 AUTOSOFT XC UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1001680 6004944

Patients

Seq Age Sex Outcome Treatment
1 10 YR Male