FDA Adverse Event Malfunction Summary report: N

ISITE PACS

MDR report key: 1984731 · Received January 27, 2011

Report

Report Number
2954704-2011-00002
Event Type
Malfunction
Date Received
January 27, 2011
Date of Event
December 9, 2010
Report Date
January 27, 2011
Manufacturer
PHILIPS HEALTHCARE INFORMATICS, INC.
Product Code
LLZ
PMA / PMN Number
K063267
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

VOLUME VISION (VIEW FORUM) IS A PART OF THE ISITE PACS SOFTWARE PACKAGE WHICH IS USED WITH GENERAL COMPUTING HARDWARE TO ACQUIRE, STORE, DISTRIBUTE, PROCESS AND DISPLAY IMAGES AND ASSOCIATED DATA THROUGHOUT A CLINICAL ENVIRONMENT. THE MALFUNCTION IS CURRENTLY UNDER DEFECT INVESTIGATION.

Description of Event or Problem · 1

WHEN PERFORMING MEASUREMENTS ON A VOLUME VISION IMAGE, CLINICIANS ARE ABLE TO TAKE SNAPSHOTS OF THEIR MEASUREMENTS IN A CAPTURE WINDOW. IF THE CAPTURE WINDOW IS NOT SAVED AND MEASUREMENT CHANGES ARE MADE TO THE ORIGINAL IMAGE, THE MEASUREMENT VALUES IN THE CAPTURE WINDOW WILL BE UPDATED. HOWEVER, THE ASSOCIATED GRAPHICS IN THE CAPTURE WINDOW WILL NOT BE UPDATED. THIS PROBLEM WAS INITIALLY DISCOVERED BY THE PRODUCT DEVELOPMENT TEAM WHILE PERFORMING TESTING OF THE NEXT VERSION OF VOLUME VISION 8.2. WHEN INVESTIGATED, IT WAS DETERMINED TO ALSO BE PRESENTED IN THE CURRENT DISTRIBUTED PRODUCT (VOLUME VISION 7.4) IT WAS BROUGHT TO THE MFR'S ATTENTION ON (B)(4), 2010. IT WAS LATER REPORTED IN A CUSTOMER COMPLAINT; COMPLAINT WAS RECEIVED ON (B)(6), 2011.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ISITE PACS ISITE PACS LLZ PHILIPS HEALTHCARE INFORMATICS, INC.

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention