FDA Adverse Event
Injury
Summary report: N
GLOBAL SHD ANGLED DRIVER
MDR report key: 1984578
·
Received January 28, 2011
Report
- Report Number
- 1818910-2011-00644
- Event Type
- Injury
- Date Received
- January 28, 2011
- Date of Event
- December 29, 2010
- Report Date
- December 29, 2010
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- LXH
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
Description of Event or Problem · 1
DURING A REAM AND RUN PROCEDURE, THE REAMER HANDLE LOCKED UP, CAUSING IT NOT TO SPIN THE REAMER ATTACHED TO IT. AS THE SURGEON WAS TORQUING ON IT TO MAKE IT SPIN, IT CAUSED SOME DEBRIS TO ESCAPE FROM THE HANDLE, WHICH WAS FOUND ON POST-OP XRAYS IN THE PT'S ARM. THE PROBLEM WITH THE REAMER HANDLE ALSO CAUSED A 20-MINUTE SURGICAL DELAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GLOBAL SHD ANGLED DRIVER | 87LXH | LXH | DEPUY ORTHOPAEDICS, INC. | NA | A0805A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |