FDA Adverse Event Injury Summary report: N

ACUSON SEQUOIA

MDR report key: 19845536 · Received July 26, 2024

Report

Report Number
3023245-2024-00048
Event Type
Injury
Date Received
July 26, 2024
Date of Event
November 10, 2022
Report Date
January 29, 2025
Manufacturer
SIEMENS HEALTHINEERS
Product Code
IYN
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

REFERENCE# (B)(4).

Additional Manufacturer Narrative · 0

THE INITIAL MDR (MFR# 3023245-2024-00048) WAS SUBMITTED FOLLOWING A RETROSPECTIVE REVIEW PERFORMED BY SIEMENS. THIS REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION RELATED TO THE INVESTIGATION RESULTS. IN THIS CASE, THE CUSTOMER REPORTED THAT THERE WAS A CRACK IN THE PLASTIC NEAR THE FRONT HANDLES WHICH LED TO A MINOR CUT ON THE SONOGRAPHER'S FINGER ON THIS IN WHILE PRESSING THE LOCK RELEASE. IT WAS CONFIRMED THAT THERE WAS NO MEDICAL INTERVENTION NEEDED FOR THE CUT. THE COVER UNDER FRONT HANDLE ASSY WAS REPLACED AND THE ISSUE HAS NOT RECURRED. A SEARCH WAS PERFORMED IN THE SERVICE AND COMPLAINTS DATABASE (GSMS AND SMARTSOLVE) AND NO OTHER RELATED ISSUES WERE FOUND FOR THIS SYSTEM. IT COULD NOT BE DETERMINED HOW THE PLASTIC WAS BROKEN BASED ON AVAILABLE INFORMATION. NO TRENDS WERE FOUND FOR THIS ISSUE. REFERENCE# (B)(4).

Description of Event or Problem · 0

THERE IS A CRACKED SMALL PLASTIC PANEL UNDERNEATH THE FRONT HANDLES OF ONE OF OUR SEQUOIA ULTRASOUND MACHINES AND ONE OF OUR SONOGRAPHER RECENTLY CUT HIS FINGER ON THIS IN WHILE PRESSING THE LOCK RELEASE BUTTONS NEAR IT. THIS MDR IS BEING SUBMITTED FOLLOWING A RETROSPECTIVE REVIEW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459942 ACUSON SEQUOIA ULTRASOUND DEVICE IYN SIEMENS HEALTHINEERS 11148775

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown