FDA Adverse Event Malfunction Summary report: N

KII SLEEVE

MDR report key: 19845085 · Received July 26, 2024

Report

Report Number
2027111-2024-00762
Event Type
Malfunction
Date Received
July 26, 2024
Date of Event
June 27, 2024
Report Date
February 4, 2025
Manufacturer
APPLIED MEDICAL RESOURCES
Product Code
GCJ
UDI-DI
00607915124267
PMA / PMN Number
K060096
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KU
Reporter Occupation
117

Narratives

Additional Manufacturer Narrative · 0

THE EVENT UNIT IS ANTICIPATED TO RETURN TO APPLIED MEDICAL FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SENT UPON COMPLETION OF INVESTIGATION.

Additional Manufacturer Narrative · 0

THE EVENT UNIT WAS RETURNED TO APPLIED MEDICAL FOR EVALUATION. VISUAL INSPECTION CONFIRMED THE COMPLAINANT¿S EXPERIENCE OF A SEVERED CANNULA WING TAB WITHIN THE CANNULA SHAFT. BASED ON THE CONDITION OF THE RETURNED UNIT, THE SEVERED WING TAB DID NOT ORIGINATE FROM THE RETURNED CANNULA AS NO DAMAGES WERE OBSERVED ON THE RETURNED CANNULA¿S WING TAB. IT IS POSSIBLE THE WING TAB FOUND WITHIN THE CANNULA SHAFT BECAME LODGED WITHIN THE CANNULA SOMETIME DURING THE MANUFACTURING PROCESS. APPLIED MEDICAL HAS PERFORMED A HISTORICAL TREND ANALYSIS AND REVIEW OF PRODUCTION RECORDS AND NO RELEVANT DOCUMENTATIONS WERE IDENTIFIED. [CORRECTION] SECTIONS D1 AND D4 (PRIMARY UNIQUE DEVICE IDENTIFICATION NUMBER AND ADDITIONAL UNIQUE DEVICE IDENTIFICATION NUMBER) HAVE BEEN UPDATED.

Description of Event or Problem · 0

PROCEDURE PERFORMED: LAP CHOLECYSTECTOMY. EVENT DESCRIPTION: "TROCAR 11 MM Z-THREAD WAS BEING USED IN A LAP CHOLE CASE. THE SURGEON WAS TRYING TO INSERT THE INTRODUCER THROUGH THE SLEEVE WHEN HE NOTICED A SMALL PLASTIC PART STUCK IN THE MIDDLE OF THE SLEEVE. HE WAS NOT ABLE TO REMOVE IT, SO ANOTHER SLEEVE WAS OPENED. NO CLINICAL IMPACT ON THE PATIENT". INTERVENTION: A NEW SLEEVE WAS OPENED PATIENT STATUS: NO CLINICAL IMPACT ON THE PATIENT.

Description of Event or Problem · 0

PROCEDURE PERFORMED: LAP CHOLECYSTECTOMY. EVENT DESCRIPTION: "TROCAR 11 MM Z-THREAD WAS BEING USED IN A LAP CHOLE CASE. THE SURGEON WAS TRYING TO INSERT THE INTRODUCER THROUGH THE SLEEVE WHEN HE NOTICED A SMALL PLASTIC PART STUCK IN THE MIDDLE OF THE SLEEVE. HE WAS NOT ABLE TO REMOVE IT, SO ANOTHER SLEEVE WAS OPENED. NO CLINICAL IMPACT ON THE PATIENT". INTERVENTION: A NEW SLEEVE WAS OPENED. PATIENT STATUS: NO CLINICAL IMPACT ON THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
921979 KII SLEEVE LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ APPLIED MEDICAL RESOURCES CTS12 UNK 00607915124267

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown