ACUSON SC2000 ULTRASOUND SYSTEM

MDR report key: 19844796 · Received July 26, 2024

Report

Report Number
3023245-2024-00051
Event Type
Malfunction
Date Received
July 26, 2024
Date of Event
March 14, 2022
Report Date
January 29, 2025
Manufacturer
SIEMENS HEALTHINEERS
Product Code
IYN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS COMPLAINT WAS CREATED AS A RESULT OF THE RETROSPECTIVE REVIEW PER EQMS-CAPA-001508 IMPLEMENTATION ACTION. REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

THE INITIAL MDR (MFR# 3023245-2024-00051) WAS SUBMITTED FOLLOWING A RETROSPECTIVE REVIEW PERFORMED BY SIEMENS. THIS REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION RELATED TO THE INVESTIGATION RESULTS. REVIEW OF THE SERVICE TICKET FOUND THAT SERVICE REVIEWED THE LOG FILES AND FOUND THE SYSTEM HANG WITH A RAD ERROR ON 3/14. ALSO, THERE WERE SOME SOFTWARE ERRORS IN THE LOGS. CSE RELOADED SOFTWARE AND REPLACED THE POWER SUPPLY MODULE. THE REPLACEMENT FIXED THE HANG ISSUE, BUT THE UNIT WOULD NOT BOOT. AFTER TROUBLESHOOTING, THE MBM250, HDD WERE REPLACED, AND THE SW WAS RE-INSTALLED TO GET THE SYSTEM WORKING. THE ISSUE DID NOT RECUR. NO TRENDS WERE FOUND FOR THIS ISSUE; THUS, NO FURTHER ACTION IS REQUIRED. REFERENCE# (B)(4).

Description of Event or Problem · 0

RETROSPECTIVE REVIEW - ACUSON SC2000 UNIT HUNG DURING STRESS ECHO EXAM. SYSTEM HUNG ON SHUTDOWN AFTER PERFORMING DIAGNOSTICS. NO INFORMATION ON HOW EXAM COMPLETED AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1587536 ACUSON SC2000 ULTRASOUND SYSTEM ULTRASOUND DEVICE IYN SIEMENS HEALTHINEERS 10433816

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization