ENDOPATH** XCEL* TROCAR
Report
- Report Number
- 3005075853-2011-00566
- Event Type
- Malfunction
- Date Received
- February 10, 2011
- Date of Event
- January 12, 2011
- Report Date
- January 13, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED AND THE LENS OF THE OBTURATOR WAS NOTED TO BE CRACKED. A POTENTIAL CAUSE OF THIS CONDITION IS THE REPEATED USE OF EO AS A STERILIZATION AGENT. SINGLE-USE TROCARS ARE NOT TO BE REUSED, REPROCESSED OR RE-STERILIZED. IN ADDITION, THE TIP OF THE SLEEVE WAS NOTED TO BE DAMAGED. ONE POSSIBLE CAUSE FOR THIS TYPE OF DAMAGE MAY BE INTERACTION WITH AN ENERGIZED DEVICE USED DURING THE PROCEDURE. CAUTION SHOULD BE TAKEN TO AVOID CONTACT BETWEEN AN ENERGIZED DEVICE AND THE TROCAR DURING THE SURGICAL PROCEDURE; HOWEVER, THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A ROBOTIC NEPHRECTOMY PROCEDURE, A PIECE OF THE TROCAR TIP WAS MISSING. NO OTHER DETAILS OF THE EVENT ARE KNOWN. THERE WAS NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH** XCEL* TROCAR | LAPAROSCOPE, GENERAL AND PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |