FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL* TROCAR

MDR report key: 1984191 · Received February 10, 2011

Report

Report Number
3005075853-2011-00566
Event Type
Malfunction
Date Received
February 10, 2011
Date of Event
January 12, 2011
Report Date
January 13, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED AND THE LENS OF THE OBTURATOR WAS NOTED TO BE CRACKED. A POTENTIAL CAUSE OF THIS CONDITION IS THE REPEATED USE OF EO AS A STERILIZATION AGENT. SINGLE-USE TROCARS ARE NOT TO BE REUSED, REPROCESSED OR RE-STERILIZED. IN ADDITION, THE TIP OF THE SLEEVE WAS NOTED TO BE DAMAGED. ONE POSSIBLE CAUSE FOR THIS TYPE OF DAMAGE MAY BE INTERACTION WITH AN ENERGIZED DEVICE USED DURING THE PROCEDURE. CAUTION SHOULD BE TAKEN TO AVOID CONTACT BETWEEN AN ENERGIZED DEVICE AND THE TROCAR DURING THE SURGICAL PROCEDURE; HOWEVER, THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A ROBOTIC NEPHRECTOMY PROCEDURE, A PIECE OF THE TROCAR TIP WAS MISSING. NO OTHER DETAILS OF THE EVENT ARE KNOWN. THERE WAS NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL* TROCAR LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1