FDA Adverse Event Death Summary report: N

ENRHYTHM DR

MDR report key: 1984188 · Received February 10, 2011

Report

Report Number
2647346-2011-00095
Event Type
Death
Date Received
February 10, 2011
Date of Event
December 24, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE. IT WAS NOTED VISUAL ANALYSIS ONLY WAS PERFORMED. (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE. IT WAS NOTED VISUAL ANALYSIS ONLY WAS PERFORMED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE. IT WAS NOTED VISUAL ANALYSIS ONLY WAS PERFORMED. (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE. IT WAS NOTED VISUAL ANALYSIS ONLY WAS PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT ARRIVED IN CARDIAC ARREST TO THE MEDICAL CENTER FIVE WEEKS AFTER PACEMAKER REPLACEMENT. FOLLOW UP INFORMATION OBTAINED INDICATED THAT THE PATIENT WAS FOUND PULSELESS AND APNEIC. RESUSCITATIVE EFFORTS WERE STARTED BY EMERGENCY MEDICAL SERVICES AND TRANSPORTED PATIENT TO THE EMERGENCY ROOM. RESUSCITATION EFFORTS CONTINUED IN THE EMERGENCY ROOM, BUT WERE UNSUCCESSFUL. NO SHOCK WAS ADVISED DURING EMERGENCY ROOM EFFORTS AND A FASTING BLOOD SUGAR OF 17 WAS OBTAINED. THE CAUSE OF DEATH HAS BEEN REQUESTED, BUT NOT RECEIVED.

Description of Event or Problem · 1

ASKU

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 59 YR Death| H| L| R (B)(4) TISSUE VALVE| (B)(4) TISSUE VALVE| (B)(4) TISSUE VALVE