FDA Adverse Event Malfunction Summary report: N

AFFINITY

MDR report key: 1984079 · Received January 14, 2011

Report

Report Number
1824206-2011-00201
Event Type
Malfunction
Date Received
January 14, 2011
Date of Event
December 21, 2010
Report Date
December 21, 2010
Manufacturer
HILL-ROM INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT THE GAS SPRINGS WERE FROZEN AND COULD NOT BE MOVED. THE TECHNICIAN REPLACED THE GAS SPRINGS AND THE TRENDELENBURG FUNCTIONED PROPERLY.

Description of Event or Problem · 1

TECHNICIAN ALLEGED THAT THE BED COULD NOT BE PUT INTO TRENDELENBURG. NO ALLEGED INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BIRTHING BED HDD HILL-ROM INC. 3600

Patients

Seq Age Sex Outcome Treatment
1