FDA Adverse Event Malfunction Summary report: N

ELECSYS ANTI-TPO

MDR report key: 19840644 · Received July 26, 2024

Report

Report Number
1823260-2024-02177
Event Type
Malfunction
Date Received
July 26, 2024
Date of Event
June 27, 2024
Report Date
September 12, 2024
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JZO
UDI-DI
04015630929702
PMA / PMN Number
K051890
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE ANTI-TPO REAGENT EXPIRATION DATE WAS NOT PROVIDED. THE COBAS E411 RACK SERIAL NUMBER WAS (B)(6). THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative · 0

SECTION D4 WAS UPDATED. THE CALIBRATION WAS NOT PERFORMED DAILY AS RECOMMENDED. THE PRODUCT LABELING STATES: "RENEWED CALIBRATION ON ALL ANALYZERS: DAILY: WHEN USING THE SAME REAGENT KIT ON THE ANALYZERS." THE PROVIDED QC SCREENSHOT SHOWED THAT THE QC WAS WITHIN RANGES, BUT THE MEASUREMENT DATE/TIME AND THE ACTUAL NUMERIC RESULTS WERE NOT SHOWN. THE PROVIDED PRE-ANALYTICAL SHOWED THAT THE CUSTOMER USED 13 MM DIAMETER TUBES BUT DID NOT USE THE REQUIRED RACK ADAPTERS FOR THEM. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Description of Event or Problem · 0

WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SERUM/PLASMA SAMPLE TESTED WITH ELECSYS ANTI-TPO ASSAY ON A COBAS E411 IMMUNOASSAY ANALYZER (RACK SYSTEM). INITIAL RESULT: >600.0 IU/ML (ACCOMPANIED BY A DATA FLAG AND CONSIDERED ABOVE THE MEASURING RANGE). THE CUSTOMER QUESTIONED THE INITIAL RESULT AS IT DID NOT MATCH THE PATIENT'S PREVIOUS RESULTS. THE SAMPLE WAS THEN REPEATED. 1ST REPEAT RESULT: 10.22 IU/ML OBTAINED WITH A MANUAL DILUTION OF 1:5 AND, THEREFORE, MULTIPLIED BY THE FACTOR 5 THE FINAL RESULT WOULD BE 51.1 IU/ML. 2ND REPEAT RESULT: 18.76 IU/ML (NO DILUTION). NO QUESTIONABLE RESULT WAS REPORTED OUTSIDE THE LABORATORY. THE 2ND REPEAT RESULT WAS DEEMED TO BE CORRECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1529779 ELECSYS ANTI-TPO SYSTEM, TEST, THYROID AUTOANTIBODY JZO ROCHE DIAGNOSTICS 769146 04015630929702

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown