STRAIGHT RASP HANDLES
Report
- Report Number
- 2249697-2011-00050
- Event Type
- Malfunction
- Date Received
- January 14, 2011
- Date of Event
- December 20, 2010
- Report Date
- December 20, 2010
- Manufacturer
- STRYKER ORTHOPAEDICS MAHWAH
- Product Code
- LXH
- PMA / PMN Number
- NA
- Removal / Correction Number
- 2249697-11/11/10-007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
SUMMARY OF EVAL: THE INVESTIGATION CONCLUDED THAT THE EVENT IS RELATED TO OTHER PREVIOUS EVENTS FOR THIS PRODUCT WHERE THE HANDLE FRACTURED INFERIOR TO THE STRIKE PLATE. IN SOME OF THESE EVENTS THE STRIKE PLATE THEN DISASSOCIATED FROM THE HANDLE. INVESTIGATIONS OF THESE PREVIOUS EVENTS CONCLUDED THAT THE ROOT CAUSE IS DESIGN RELATED. THE POSITION OF THE THROUGH HOLE, VERSION HOLES AND SUPERIOR LIGHTENING HOLE, INFERIOR TO THE STRIKE PLATE, DO NOT DISCOURAGE CRACK INITIATION AND PROPAGATION. NO SIMILAR EVENTS HAVE BEEN REPORTED FOR PRODUCTS MANUFACTURED AFTER IMPLEMENTATION OF THIS DESIGN CHANGE. IF CRACKS ARE FOUND THE HANDLES ARE TO BE RETURNED TO STRYKER ORTHOPAEDICS UNDER (B)(4).
THE DEVICES WERE INSPECTED BY THE CUSTOMER AND THE ITEM FAILED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STRAIGHT RASP HANDLES | INSTRUMENT | LXH | STRYKER ORTHOPAEDICS MAHWAH | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Required Intervention |