FDA Adverse Event
Malfunction
Summary report: N
R/R TPS ELITE LONG ANGLE ATT
MDR report key: 1983848
·
Received January 11, 2011
Report
- Report Number
- 1811755-2011-00063
- Event Type
- Malfunction
- Date Received
- January 11, 2011
- Date of Event
- December 18, 2010
- Report Date
- December 18, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DRILL ATTACHMENT WAS CONTAINED BY THE MANUFACTURER AND THE DEVICE WAS DISASSEMBLED. FOREIGN DEBRIS WAS FOUND SURROUNDING THE NOSE BEARING, PREVENTING IT FROM ROTATING FREELY. DEBRIS WITHIN BEARINGS WILL CAUSE EXCESSIVE FRICTION, WHICH CAN LEAD TO HEAD BEING GENERATED. THE DRILL ATTACHMENT COULD NOT BE REPAIRED, AND THEREFORE WAS NOT RETURNED TO THE USER FACILITY.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER, THE DRILL ATTACHMENT HEATED UP. THIS EVENT DID NOT OCCUR IN A SURGICAL ENVIRONMENT, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | R/R TPS ELITE LONG ANGLE ATT | DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |