FDA Adverse Event Malfunction Summary report: N

R/R TPS ELITE LONG ANGLE ATT

MDR report key: 1983848 · Received January 11, 2011

Report

Report Number
1811755-2011-00063
Event Type
Malfunction
Date Received
January 11, 2011
Date of Event
December 18, 2010
Report Date
December 18, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DRILL ATTACHMENT WAS CONTAINED BY THE MANUFACTURER AND THE DEVICE WAS DISASSEMBLED. FOREIGN DEBRIS WAS FOUND SURROUNDING THE NOSE BEARING, PREVENTING IT FROM ROTATING FREELY. DEBRIS WITHIN BEARINGS WILL CAUSE EXCESSIVE FRICTION, WHICH CAN LEAD TO HEAD BEING GENERATED. THE DRILL ATTACHMENT COULD NOT BE REPAIRED, AND THEREFORE WAS NOT RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER, THE DRILL ATTACHMENT HEATED UP. THIS EVENT DID NOT OCCUR IN A SURGICAL ENVIRONMENT, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 R/R TPS ELITE LONG ANGLE ATT DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK