FDA Adverse Event Malfunction Summary report: N

R/R SINGLE TRIG PIN COLLET

MDR report key: 1983785 · Received January 11, 2011

Report

Report Number
1811755-2011-00064
Event Type
Malfunction
Date Received
January 11, 2011
Date of Event
December 16, 2010
Report Date
December 16, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE WIRE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PINS WERE STICKING IN THE COLLET DURING TESTING IN HOUSE. THERE WAS NO PT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 R/R SINGLE TRIG PIN COLLET INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR KIJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK