FDA Adverse Event Malfunction Summary report: N

OLYMPUS RESECTION SHEATH

MDR report key: 1983694 · Received February 3, 2011

Report

Report Number
9610773-2011-00004
Event Type
Malfunction
Date Received
February 3, 2011
Date of Event
January 5, 2011
Report Date
January 5, 2011
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
FAS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

OLYMPUS FOLLOWED UP WITH THE USER FACILITY TO OBTAIN ADDITIONAL INFO AND WAS INFORMED OF THE FOLLOWING INFO: THE DEVICE WAS REPORTED TO HAVE BEEN USED IN THE PROCEDURE FOR ABOUT FIVE TO TEN MINUTES WHEN THE CERAMIC TIP OF THE RESECTION SHEATH REPORTEDLY BROKE AND FELL INTO THE PT. THE BROKEN CERAMIC TIP WAS IMMEDIATELY RETRIEVED USING GRASPER FORCEPS. THE PROCEDURE WAS COMPLETED USING A DIFFERENT DEVICE. THE STAFF REPORTED THAT AN X-RAY WAS PERFORMED SUBSEQUENT TO THE REPORTED EVENT AND REVEALED NO EVIDENCE OF DEVICE FRAGMENTS INSIDE THE PT. THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS FOR EVAL. THE CERAMIC BEAK ON THE DISTAL END OF THE UNIT WAS HEAVILY DAMAGED. A PORTION OF THE CERAMIC BEAK WAS BROKEN OFF AND NOT RETURNED, AND THERE WAS EVIDENCE OF THERMAL DAMAGE AT THE EDGE OF A PORTION OF THE FRACTURE JUNCTION. BASED UPON THE FINDINGS OF THIS INVESTIGATION, IT APPEARS THE CAUSE OF THIS PHENOMENON WAS RELATED TO PHYSICAL DAMAGE. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT (MDR) IN AN ABUNDANCE OF CAUTION. (B)(4).

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT DURING A TRANSURETHRAL RESECTION OF PROSTATE (TURP) PROCEDURE A PORTION OF THE CERAMIC TIP OF A RESECTION SHEATH BROKE OFF AND FELL INSIDE THE PT. THE BROKEN PIECE WAS RETRIEVED USING AN UNK MODEL OF GRASPER FORCEPS. THE PROCEDURE WAS COMPLETED AND THERE WAS NO PT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OLYMPUS RESECTION SHEATH RESECTION SHEATH FAS OLYMPUS WINTER & IBE GMBH A22042A UNK

Patients

Seq Age Sex Outcome Treatment
1 78 YR