FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 1983690 · Received February 3, 2011

Report

Report Number
3002158293-2011-00117
Event Type
Malfunction
Date Received
February 3, 2011
Date of Event
December 23, 2010
Report Date
February 1, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BELT WON'T STAY LATCHED TO MONITOR) HAS BEEN CONFIRMED. UPON EVAL, THE TRUNK CABLE CONNECTOR WAS BROKEN. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE BROKEN CONNECTOR. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

ZOLL CUSTOMER SUPPORT CONTACTED A (B)(6) MALE PT IN REGARDING TO AN UNRELATED ISSUE WITH HIS BELT. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT. UPON SERVICING OF BELT (B)(4), IT WAS FOUND THAT THE BELT WOULD NOT STAY SECURELY LATCHED TO THE MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 68 YR