FDA Adverse Event Malfunction Summary report: N

DIMENSION® CLINICAL CHEMISTRY SYSTEM

MDR report key: 1983518 · Received February 10, 2011

Report

Report Number
1226181-2011-00008
Event Type
Malfunction
Date Received
February 10, 2011
Date of Event
January 26, 2011
Report Date
January 26, 2011
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD
Product Code
MMI
PMA / PMN Number
K944093
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS R2 PROBE MALFUNCTION. SIEMENS HEALTHCARE DIAGNOSTICS TECHNICAL SUPPORT CENTER DIRECTED THE CUSTOMER TO REPLACEMENT OF THE R2 PROBE AND OTHER MAINTENANCE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED. THE PATIENT WAS ADMITTED. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION® CLINICAL CHEMISTRY SYSTEM CLINICAL CHEMISTRY SYSTEM MMI SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD RXLMAXRHQ

Patients

Seq Age Sex Outcome Treatment
1