ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2011-00540
- Event Type
- Malfunction
- Date Received
- February 10, 2011
- Date of Event
- January 31, 2011
- Report Date
- January 31, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND NOT RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED. THE DEVICE WAS ACTIVATED WITH THE GENERATOR AND AN ERROR CODE 5 WAS DISPLAYED. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN AN ERROR CODE 5 OR BLADE "LOCKOUT" LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE JAW BROKE OFF AND FELL INTO THE PATIENT. THE SURGEON RETRIEVED THE PIECE FROM THE PATIENT WITH NO PATIENT CONSEQUENCE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | G9LP1Y |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |