FDA Adverse Event
Malfunction
Summary report: N
DIMENSION® CLINICAL CHEMISTRY SYSTEM
MDR report key: 1983507
·
Received February 10, 2011
Report
- Report Number
- 1226181-2011-00013
- Event Type
- Malfunction
- Date Received
- February 10, 2011
- Date of Event
- January 28, 2011
- Report Date
- January 28, 2011
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD
- Product Code
- MMI
- PMA / PMN Number
- K944093
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS DUE TO A CONTAMINATION FROM HM, SAMPLE PROBE, OR DRAIN. A SIEMENS HEALTHCARE TECHNICAL SERVICE CENTER ASSOCIATE DIRECTED THE CUSTOMER TO MAINTENANCE PROCEDURES. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
Description of Event or Problem · 1
A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.