ETS FLEX ARTICNG LNR
Report
- Report Number
- 3005075853-2011-00542
- Event Type
- Malfunction
- Date Received
- February 10, 2011
- Date of Event
- July 10, 2010
- Report Date
- September 24, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4). FIRING TRIGGER TEETH, SPRING CARTRIDGE LOCKOUT TAB. THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AS THE FIRING TRIGGER WAS JAMMED HALFWAY. A RELOAD WAS LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED PARTIALLY FIRED AND WITH CARTRIDGE LOCKOUT SPRING DAMAGED. THE DAMAGE TO THE RELOAD LOCKOUT SPRING IS CONSISTENT WITH DAMAGE OBSERVED WHEN THE FIRING CYCLE IS STARTED, INTERRUPTED, RELEASED, AND RESTARTED. WHEN FIRING THE DEVICE MAKE SURE THAT THE FIRING STROKE IS COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR MORE INFORMATION. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A PANCREAS PSEUDO CYST PROCEDURE, THE STAPLER DID NOT WORK. THE STAPLER DID NOT STAPLE AND DID NOT CUT. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETS FLEX ARTICNG LNR | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | G4R524 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |