PENTAX
Report
- Report Number
- 9610877-2024-55766
- Event Type
- Malfunction
- Date Received
- July 26, 2024
- Date of Event
- July 17, 2024
- Report Date
- July 26, 2024
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- CAL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OCULAR FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE OCULAR. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE CONTROL BODY FLUID DAMAGE, THE FORWARD BODY FRAME FLUID DAMAGE, THE CFB SCREEN COVER GLASS FLUID DAMAGE, THE CFB (COHERENT FIBER BUNDLE) FLUID DAMAGE, THE U/D KNOB BROKEN, THE INSERTION FLEXIBLE TUBE PERFORATED, AND THE INSERTION FLEXIBLE TUBE CRUSHED; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. THIS DEFECT OCCURRED NOT DURING PROCEDURE. BASED ON THE TECHNICAL REPORT "HR-RPT-0586 (IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS NOT DURING PROCEDURE. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (FLUID DAMAGE).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1733709 | PENTAX | FIBEROPTICINTUBATIONSCOPE 1.2C 3.0TP ROP | CAL | HOYA CORPORATION PENTAX TOKYO OFFICE | FI-9RBS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |