TRUSTEEL
Report
- Report Number
- 3003442380-2024-16542
- Event Type
- Malfunction
- Date Received
- July 25, 2024
- Date of Event
- June 18, 2024
- Report Date
- November 12, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244018426
- PMA / PMN Number
- K041545
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR 1921387 - MDR 3003442380-2024-16542 - DEVICE 5 OF 8.
SUPPLEMENTAL REPORT 01 - (B)(4) - MDR. CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. PER REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DHR REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DHR REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6003366, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6003366 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 94 AND PACKAGED IN THE MACHINE MULTIVAC 10 ON 26-SEP-2023, WITH A TOTAL OF (B)(4) UNITS. TUBE GLUING. THE SUB-ASSEMBLY, TUBE GLUING OF THE LOT 3J02661 WAS MANUFACTURED ACCORDING TO THE WI VERSION 55 AND MANUFACTURED IN THE MACHINE SC05-SC06, ON 25-SEP-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE; NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN GERMANY. IT WAS REPORTED THAT PATIENT FACED 8 INFUSIONS SET LEAKAGE BETWEEN NEEDLE AND PATCH EVENT ON 18-JUN-2024. INFUSION SET HAS BEEN USED FOR 12 HOURS. THE BLOOD GLUCOSE LEVEL WAS 255 AND 300MG/DL. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 544333 | TRUSTEEL | UNO CONTACT DETACH G29 60/6TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1002833 | 6003366 | 05705244018426 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |