FDA Adverse Event Malfunction Summary report: N

INSET II

MDR report key: 19832920 · Received July 25, 2024

Report

Report Number
3003442380-2024-16391
Event Type
Malfunction
Date Received
July 25, 2024
Date of Event
June 20, 2024
Report Date
November 13, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018129
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED BRAND NAME UNDER D1 AND EXPIRATION DATE UNDER D4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DEVICE HISTORY RECORD (DHR) REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DEVICE HISTORY RECORD (DHR) REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6003497 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6003497 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 108 AND MANUFACTURED IN THE LINE # 1 ON 01/OCT/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. ON 20-JUNE-2024, IT WAS REPORTED BY THE PATIENT THAT HE RECEIVES AN ALARM. IN TROUBLESHOOTING BLOCKAGE WAS FOUND AT THE SITE. NO FURTHER INFORMATION WAS AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
544836 INSET II UNO INSET II 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1002817 6003497 05705244018129

Patients

Seq Age Sex Outcome Treatment
1 40 YR Female