FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1983280 · Received February 2, 2011

Report

Report Number
2531779-2011-00677
Event Type
Malfunction
Date Received
February 2, 2011
Report Date
January 6, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE DOWN ARROW AND OK BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE DOWN ARROW AND OK KEY CONTACTS.

Description of Event or Problem · 1

THE PT CLAIMED THAT THE PRIME STEP DID NOT IMMEDIATELY STOP WHEN THE OK BUTTON WAS PRESSED. SHE ALSO CLAIMED THAT THE KEYPAD BUTTONS FEEL LIKE THEY ARE STUCK DOWN. THE PATIENT DENIED EXPOSING THE PUMP TO MOISTURE AND INDICATED THAT THERE WAS NO EVIDENCE THAT THE KEYPAD WAS PEELING. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1