FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1983274 · Received February 2, 2011

Report

Report Number
2531779-2011-00611
Event Type
Malfunction
Date Received
February 2, 2011
Date of Event
January 7, 2011
Report Date
January 7, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Removal / Correction Number
2531779-03/24/2010/003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALED LOSS OF PRIME WARNINGS ASSOCIATED WITH ZERO FORCE. THE PUMP WAS REWOUND, LOADED, AND PRIMED SUCCESSFULLY WITH NO ALARMS OCCURRING.

Description of Event or Problem · 1

THE PT STATED THAT THE PUMP DISPENSED INSULIN FROM THE CARTRIDGE DURING THE LOAD STEP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 68 YR