FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1983272 · Received February 2, 2011

Report

Report Number
2531779-2011-00674
Event Type
Malfunction
Date Received
February 2, 2011
Report Date
January 6, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE OK BUTTON WAS INTERMITTENTLY RESPONDING TO USER INPUTS AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE DOWN ARROW KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UP/DOWN ARROW, CONTRAST, AND OK BUTTONS WERE INCREASINGLY NOT ACTIVATING THE DESIRED PUMP FUNCTIONS. THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE AND THE KEYPAD IS NOT PEELING; HOWEVER, THE BUTTONS FEEL FLAT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment