FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1983272
·
Received February 2, 2011
Report
- Report Number
- 2531779-2011-00674
- Event Type
- Malfunction
- Date Received
- February 2, 2011
- Report Date
- January 6, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE OK BUTTON WAS INTERMITTENTLY RESPONDING TO USER INPUTS AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE DOWN ARROW KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE UP/DOWN ARROW, CONTRAST, AND OK BUTTONS WERE INCREASINGLY NOT ACTIVATING THE DESIRED PUMP FUNCTIONS. THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE AND THE KEYPAD IS NOT PEELING; HOWEVER, THE BUTTONS FEEL FLAT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |