FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1983266 · Received February 2, 2011

Report

Report Number
2531779-2011-00684
Event Type
Malfunction
Date Received
February 2, 2011
Report Date
January 6, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. INVESTIGATION REVEALED THAT THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE UP/DOWN/OK/CONTRAST ARROW BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS AND REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, THE UP/DOWN KEY CONTACTS WERE MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED, THE UP/DOWN/OK KEY CONTACTS BECAME INVERTED WHEN PRESSED, AND THERE WAS EVIDENCE OF CONTAMINATION UNDER ALL KEYPAD KEY CONTACTS.

Description of Event or Problem · 1

THE PT'S MOM CLAIMED THAT THE UP/DOWN/OK BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1