FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1983266
·
Received February 2, 2011
Report
- Report Number
- 2531779-2011-00684
- Event Type
- Malfunction
- Date Received
- February 2, 2011
- Report Date
- January 6, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. INVESTIGATION REVEALED THAT THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE UP/DOWN/OK/CONTRAST ARROW BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS AND REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, THE UP/DOWN KEY CONTACTS WERE MISALIGNED, THE CONTRAST KEY CONTACT WAS INVERTED, THE UP/DOWN/OK KEY CONTACTS BECAME INVERTED WHEN PRESSED, AND THERE WAS EVIDENCE OF CONTAMINATION UNDER ALL KEYPAD KEY CONTACTS.
Description of Event or Problem · 1
THE PT'S MOM CLAIMED THAT THE UP/DOWN/OK BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING |