FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 1983260
·
Received February 2, 2011
Report
- Report Number
- 2531779-2011-00646
- Event Type
- Malfunction
- Date Received
- February 2, 2011
- Date of Event
- January 7, 2011
- Report Date
- January 7, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED EVIDENCE OF MOISTURE INGRESS ON THE PCB COMPONENTS. VISUAL INSPECTION DID NOT REVEAL ANY PUMP DEFECT. REVIEW OF THE PUMP DOWNLOAD HISTORY VERIFIED ONE POWER ON RESET EVENT AFTER A REWIND COMMAND. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT ANY POWER LOSS OR REBOOTING DUPLICATED.
Description of Event or Problem · 1
THE REPORTER ALLEGED THAT THE PUMP KEPT SHOWING THE VERIFY SCREEN AND REBOOTED WITHOUT USER INTERVENTION. SHE DENIED THAT THE PUMP HAD ANY CRACKS OR DAMAGE TO THE BATTERY COMPARTMENT AND THE BATTERY CAP WAS SECURED PROPERLY WITH A COIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |