FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1983260 · Received February 2, 2011

Report

Report Number
2531779-2011-00646
Event Type
Malfunction
Date Received
February 2, 2011
Date of Event
January 7, 2011
Report Date
January 7, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED EVIDENCE OF MOISTURE INGRESS ON THE PCB COMPONENTS. VISUAL INSPECTION DID NOT REVEAL ANY PUMP DEFECT. REVIEW OF THE PUMP DOWNLOAD HISTORY VERIFIED ONE POWER ON RESET EVENT AFTER A REWIND COMMAND. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT ANY POWER LOSS OR REBOOTING DUPLICATED.

Description of Event or Problem · 1

THE REPORTER ALLEGED THAT THE PUMP KEPT SHOWING THE VERIFY SCREEN AND REBOOTED WITHOUT USER INTERVENTION. SHE DENIED THAT THE PUMP HAD ANY CRACKS OR DAMAGE TO THE BATTERY COMPARTMENT AND THE BATTERY CAP WAS SECURED PROPERLY WITH A COIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 18 YR