FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1983190 · Received February 2, 2011

Report

Report Number
2531779-2011-00581
Event Type
Malfunction
Date Received
February 2, 2011
Report Date
January 3, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILLED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT, ALONG WITH HER PARENTS, CONTACTED ANIMAS ALLEGING THAT SHE WAS IN THE MIDDLE OF DELIVERING A BOLUS AND IT WAS CANCELLED WITHOUT HER PRESSING ANY BUTTONS. ACCORDING TO THE PUMP'S BOLUS HISTORY, 0.90 UNITS OF 6.90 UNITS WERE DELIVERED AND IT WAS NOTED TO BE CANCELLED. THE PT REPORTEDLY WAS ABLE TO DELIVER THE REMAINING AMOUNT OF INSULIN WITH NO CANCELLATIONS. THE KEYPAD REPORTEDLY IS INTACT AND THE BUTTONS ALL FEEL NORMAL AND BOUNCE/SPRING BACK. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. HOWEVER, THIS COMPLAINT IS BEING REPORTED BECAUSE THE DELIVERY OF A BOLUS REPORTEDLY WAS CANCELLED WITH NO USER INTERVENTION. A REPLACEMENT PUMP WAS SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1