ECHELON*FLEX60
Report
- Report Number
- 1527736-2011-00040
- Event Type
- Malfunction
- Date Received
- February 10, 2011
- Date of Event
- January 17, 2011
- Report Date
- January 18, 2011
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.
(B)(4). THE ANALYSIS FOUND THAT THE EC60A DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH AN ECR60G RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED PARTIALLY FIRED. A PARTIAL FIRE CAN OCCUR IF THE FIRING SEQUENCE IS INTERRUPTED. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE CUT WAS NOTED TO BE JAGGED DUE TO A DAMAGED KNIFE. ONE POSSIBLE CAUSE FOR THIS TYPE OF DAMAGE TO THE KNIFE IS WHEN THE DEVICE IS FIRED OVER AN ALREADY EXISTING STAPLE LINE, HARD OBJECT OR THICKER TISSUE THAN INDICATED. REPEATEDLY FIRING ACROSS EXISTING STAPLE LINES CAN ALSO REDUCE THE ABILITY TO CUT CLEANLY. TO MITIGATE THE POTENTIAL FOR STAPLES GETTING INTO THE CARTRIDGE AND INTERFERING WITH THE KNIFE PATH DURING DEVICE FIRING, PRIOR TO RELOADING THE DEVICE, RINSE THE ANVIL AND CARTRIDGE JAW IN STERILE SOLUTION AND THEN WIPE THE ANVIL AND CARTRIDGE JAW TO CLEAN ANY FORMED BUT UNUSED STAPLES FROM THE DEVICE. ADDITIONALLY, PROPER CARE SHOULD BE TAKEN WHEN PLACING THE DEVICE ON THE TISSUE TO BE STAPLED, TO ENSURE THAT NO HARD OBSTRUCTION SUCH AS A CLIP IS INCLUDED WITH THE TISSUE INSIDE THE JAWS. AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE, THE DEVICE WAS FIRED FOR THE FOURTH TIME AND THE STAPLE LINE HAD MALFORMED STAPLES. THEY HAD TO OVERSEW THE STAPLE LINE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ECHELON*FLEX60 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO SURGERY, INC (CINCINNATI) | UNK | G4UG6P |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |