COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2011-00795
- Event Type
- Malfunction
- Date Received
- February 9, 2011
- Date of Event
- November 1, 2010
- Report Date
- November 18, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K041191
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH LOW BATTERY WAS CONFIRMED DURING PRODUCT EVALUATION BY BAXTER QUALITY ENGINEERING. IT WAS DETERMINED BY BAXTER QUALITY ENGINEERING THAT THE REPORTED CONDITION WAS RELATED TO FAILURE CODE 814:01 WHICH WAS CAUSED BY THE DEPLETION OF THE MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE THEREFORE REPLACED TO RESOLVE THIS ISSUE. REVIEW OF THE EVENT HISTORY DETERMINED THAT THE REPORTED CONDITION OCCURRED ON (B)(6) 2010 NOT ON THE REPORTED OCCURRENCE DATE OF (B)(6) 2010. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW REVEALED THE DEVICE HAS BEEN PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED A LOW BATTERY. IT IS UNKNOWN WHICH PROCESS STEP THIS EVENT OCCURRED DURING. ENGINEERING REVIEW OF THE DEVICE EVENT HISTORY CONFIRMED THIS CONDITION. IT WAS DETERMINED BY BAXTER QUALITY ENGINEERING THAT THE REPORTED CONDITION IS RELATED TO FAILURE CODE 814:01, WHICH INTERRUPTED DELIVERY. THERE WAS NO PATIENT INVOLVEMENT, AND THEREFORE NO PATIENT INJURY OR MEDICAL INTERVENTION. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |