FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1982941 · Received February 9, 2011

Report

Report Number
6000001-2011-00795
Event Type
Malfunction
Date Received
February 9, 2011
Date of Event
November 1, 2010
Report Date
November 18, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Removal / Correction Number
6000001-12/1/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH LOW BATTERY WAS CONFIRMED DURING PRODUCT EVALUATION BY BAXTER QUALITY ENGINEERING. IT WAS DETERMINED BY BAXTER QUALITY ENGINEERING THAT THE REPORTED CONDITION WAS RELATED TO FAILURE CODE 814:01 WHICH WAS CAUSED BY THE DEPLETION OF THE MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE THEREFORE REPLACED TO RESOLVE THIS ISSUE. REVIEW OF THE EVENT HISTORY DETERMINED THAT THE REPORTED CONDITION OCCURRED ON (B)(6) 2010 NOT ON THE REPORTED OCCURRENCE DATE OF (B)(6) 2010. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW REVEALED THE DEVICE HAS BEEN PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED A LOW BATTERY. IT IS UNKNOWN WHICH PROCESS STEP THIS EVENT OCCURRED DURING. ENGINEERING REVIEW OF THE DEVICE EVENT HISTORY CONFIRMED THIS CONDITION. IT WAS DETERMINED BY BAXTER QUALITY ENGINEERING THAT THE REPORTED CONDITION IS RELATED TO FAILURE CODE 814:01, WHICH INTERRUPTED DELIVERY. THERE WAS NO PATIENT INVOLVEMENT, AND THEREFORE NO PATIENT INJURY OR MEDICAL INTERVENTION. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1