SELECT SECURE
Report
- Report Number
- 2649622-2011-00216
- Event Type
- Injury
- Date Received
- February 9, 2011
- Date of Event
- December 15, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P030036
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY - (B)(4)- NO ANOMALIES FOUND.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
ASKU
IT WAS REPORTED THE VENTRICULAR LEAD IMPEDANCE DROPPED FROM 540 TO 360 OHMS, BUT HAD BEEN STEADY AT 360 OHMS. LEAD THRESHOLDS ON BOTH THE ATRIAL AND VENTRICULAR LEADS HAD BEEN OVER 2.5V. THE VENTRICULAR LEAD WAS CAPPED AND REPLACED WITH A COMPETITOR LEAD. THE ATRIAL LEAD REMAINS IN USE BECAUSE SENSING FUNCTIONS WERE INTACT AND "PATIENT DOES NOT ATRIALLY PACE OFTEN." PHYSICIAN ELECTIVELY REPLACED DEVICE AS IT WAS NEARING ELECTIVE REPLACEMENT INDICATOR AND PATIENT HAD COMPLETE AV (ATRIOVENTRICULAR) BLOCK. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. LATER REPORTED BY MOTHER THAT AT EACH DEVICE CHECK, THE BATTERY SHOWED AS DEPLETING. SHE ALSO REPORTED "IT WAS MEDTRONIC'S MALFUNCTIONING LEAD" THAT WAS AN ISSUE.
IT WAS REPORTED THE VENTRICULAR LEAD IMPEDANCE DROPPED FROM 540 TO 360 OHMS, BUT HAD BEEN STEADY AT 360 OHMS. LEAD THRESHOLDS ON BOTH THE ATRIAL AND VENTRICULAR LEADS HAD BEEN OVER 2.5V. THE VENTRICULAR LEAD WAS CAPPED AND REPLACED WITH A COMPETITOR LEAD. THE ATRIAL LEAD REMAINS IN USE BECAUSE SENSING FUNCTIONS WERE INTACT AND "PATIENT DOES NOT ATRIALLY PACE OFTEN." NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THE VENTRICULAR LEAD IMPEDANCE DROPPED FROM 540 TO 360 OHMS, BUT HAD BEEN STEADY AT 360 OHMS. LEAD THRESHOLDS ON BOTH THE ATRIAL AND VENTRICULAR LEADS HAD BEEN OVER 2.5V. THE VENTRICULAR LEAD WAS CAPPED AND REPLACED WITH A COMPETITOR LEAD. THE ATRIAL LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SELECT SECURE | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 3830 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR | Hospitalization| L| R | (B)(4) IMPLANTABLE PULSE GENERATOR| (B)(4) IMPLANTABLE PULSE GENERATOR |