SIGMA 300 SR
Report
- Report Number
- 2647346-2011-00091
- Event Type
- Death
- Date Received
- February 9, 2011
- Date of Event
- January 24, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S2
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. IT WAS NOTED THAT PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES. (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT VISUAL ANALYSIS ONLY WAS PERFORMED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. IT WAS NOTED THAT PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES. (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT VISUAL ANALYSIS ONLY WAS PERFORMED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT VISUAL ANALYSIS ONLY WAS PERFORMED. (B)(4) NO ANOMALIES FOUND (B)(4) PROXIMAL SEGMENT.
ASKU
ASKU
THE DEVICE AND LEAD WERE RETURNED TO THE MANUFACTURER AFTER THE PATIENT'S DEATH. IT WAS REPORTED THAT THE PATIENT DIED OF ACUTE CARDIAC ARREST AND ATHEROSCLEROTIC HEART DISEASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIGMA 300 SR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | SSR303 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 89 YR | Death| O |