FDA Adverse Event Malfunction Summary report: N

ADULT DUAL-HEATED BREATHING CIRCUIT

MDR report key: 1982822 · Received February 9, 2011

Report

Report Number
9611451-2011-00062
Event Type
Malfunction
Date Received
February 9, 2011
Report Date
January 27, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THE RT200 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE INSPIRATORY AND EXPIRATORY LIMBS OF THE BREATHING CIRCUIT WERE TESTED FOR DAMAGED PINS. RESULTS: A HEATER WIRE PIN ON THE INSPIRATORY LIMB OF THE BREATHING CIRCUIT WAS SPLIT PREVENTING FULL INSERTION OF A HEATER WIRE ADAPTOR. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100309. CONCLUSION: ALL BREATHING CIRCUITS ARE TESTED FOR CONNECTIVITY AND ELECTRICAL CONTINUITY DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. IT IS POSSIBLE FOR THE USER TO BEND THE HEATER WIRE PINS DURING USE IF THE HEATER WIRE ADAPTOR IS INSERTED INTO THE HEATER WIRE PLUG ON AN ANGLE. IT IS NOT POSSIBLE FOR US TO DETERMINE WHETHER THE PINS WERE BENT DURING PRODUCTION OR BY THE USER. DAMAGED PINS DO NOT PRECLUDE VENTILATION OF THE PATIENT, BUT PREVENT THE HEATING OF THE GAS DELIVERED. (B)(4).

Description of Event or Problem · 1

A DISTRIBUTOR IN (B)(6) REPORTED THAT AN RT200 BREATHING CIRCUIT HAD A BROKEN HEATER WIRE PIN WHICH PREVENTED THE CIRCUIT FROM BEING CONNECTED TO A HEATER WIRE ADAPTOR. THE DISTRIBUTOR IDENTIFIED THE BROKEN PIN PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL-HEATED BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED RT200 100309

Patients

Seq Age Sex Outcome Treatment
1