INTEGRATED APD SET W/CASSETTE3 PRONG
Report
- Report Number
- 1423500-2011-01696
- Event Type
- Malfunction
- Date Received
- February 9, 2011
- Date of Event
- January 17, 2011
- Report Date
- January 17, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE IS NOT AVAILABLE FOR EVALUATION.A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
(B)(4). THIS COMPLAINT FOR THE SYSTEM ERROR 2240 (AIR IN LINE) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE ROOT CAUSE OF THE COMPLAINT WAS NOT DETERMINED. THE LOT NUMBER IS UNKNOWN; THEREFORE, A BATCH REVIEW CANNOT BE PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR 2240 ALARM THAT OCCURRED ON THE HOMECHOICE (HC) DURING THE INITIAL DRAIN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARMS AND ASSISTED IN TROUBLESHOOTING. THE HP WAS CONNECTED AT THE TIME OF THE CALL, AND DID NOT DISCONNECT. THE TSR EXPLAINED TO DISCARD ALL SUPPLIES AND TO REPORT ALARMS TO THE PERITONEAL DIALYSIS NURSE THE NEXT MORNING. THE HP WOULD FINISH TONIGHT'S THERAPY MANUALLY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTEGRATED APD SET W/CASSETTE3 PRONG | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | HOMECHOICE MACHINE |