FDA Adverse Event Malfunction Summary report: N

HEARTSTRING III PROXIMAL SEAL SYSTEM, 3.8 MM

MDR report key: 1982762 · Received January 6, 2011

Report

Report Number
2242352-2010-03840
Event Type
Malfunction
Date Received
January 6, 2011
Date of Event
December 7, 2010
Report Date
December 8, 2010
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: VISUAL INSPECTION REVEALED THAT THE DELIVERY TUBE WAS RETURNED SEPARATE FROM THE LOADING DEVICE. THE SEAL AND THE TENSION SPRING ASSEMBLY WERE INSIDE THE BODY OF THE LOADING DEVICE. THE GREEN SLIDE LOCK AND THE WHITE PLUNGER WERE NOT DEPRESSED ON THE DELIVERY DEVICE. THERE WAS NO EVIDENCE OF BLOOD ON THE DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT "SEAL STAYED IN THE ROLLER AND DID NOT COME OUT OF THE LOADER" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. INTERNAL FILE NUMBER - (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE SEAL STAYED IN THE ROLLER AND DID NOT COME OUT OF THE LOADER. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS. THE PRODUCT WAS RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTRING III PROXIMAL SEAL SYSTEM, 3.8 MM PROXIMAL SEAL DXC MAQUET CARDIOVASCULAR, LLC HSK-3038 25022681

Patients

Seq Age Sex Outcome Treatment
1 NA