FDA Adverse Event Malfunction Summary report: N

PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED

MDR report key: 1982739 · Received February 9, 2011

Report

Report Number
1423500-2011-01697
Event Type
Malfunction
Date Received
February 9, 2011
Date of Event
December 11, 2010
Report Date
January 19, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(6) THE DEVICE HAS BEEN RECEIVED BY BAXTER, HOWEVER, THE EVALUATION HAS NOT BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE PRODUCT ANALYSIS LAB EVALUATED THE DEVICE AND FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) DURING EVALUATION. BASED ON A REVIEW OF ALL AVAILABLE THERAPY LOG DATA, THE CAUSE OF THE IIPV WAS DETERMINED TO BE INSUFFICIENT DRAIN, ONE OR MORE CYCLES ADVANCES TO NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE IIPV. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED ON (B)(6) 2010 DURING DRAIN CYCLE 4. THE PATIENT'S DRAIN VOLUME WAS 3211ML. THE FILL VOLUME WAS 2000ML. THIS MEETS IIPV CRITERIA. PRODUCT SURVEILLANCE CONTACTED THE NURSE ON 02/09/2011. ACCORDING TO THE NURSE SHE WAS NOT AWARE OF THE IIPV EVENT. ADDITIONALLY, THE PATIENT WAS SEEN SINCE THIS EVENT AND HAS NOT REPORTED ANY OVERFILL SYMPTOMS OR ISSUES WITH THERAPY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 60 YR