FDA Adverse Event Injury Summary report: N

METRICIDE OPA PLUS

MDR report key: 1982705 · Received February 4, 2011

Report

Report Number
1221361-2011-00001
Event Type
Injury
Date Received
February 4, 2011
Date of Event
January 1, 2011
Report Date
February 3, 2011
Manufacturer
ALDEN MEDICAL LLC
Product Code
MED
PMA / PMN Number
K070627
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

END USER FACILITY USED DISINFECTANT (METRICIDE OPA PLUS) TO DISINFECT A BRONCHOSCOPE. FOLLOWING DISINFECTION, THE BRONCHOSCOPE REQUIRES A RINSING PROCESS TO REMOVE THE DISINFECTANT FROM ITS SURFACES PRIOR TO ITS USE ON A PATIENT. THE END USER HALTED THE RINSING PROCESS APPROXIMATELY 6 MINUTES BEFORE REACHING ITS COMPLETION, AND THE BRONCHOSCOPE WAS THEN USED ON A PATIENT. FOLLOWING USE, THE PATIENT EXPERIENCED PAIN AND DISCOMFORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 METRICIDE OPA PLUS MED ALDEN MEDICAL LLC NOT AVAILABLE

Patients

Seq Age Sex Outcome Treatment
1 43 YR Hospitalization