FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 1982635 · Received January 13, 2011

Report

Report Number
2531779-2011-00288
Event Type
Malfunction
Date Received
January 13, 2011
Report Date
December 16, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE UP/DOWN/OK KEY BUTTONS WERE INTERMITTENTLY RESPONDING TO USER INPUTS, THE CONTRAST BUTTON REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, THE DOWN KEY CONTACT WAS MISALIGNED, THE UP/DOWN/OK/CONTRAST KEY CONTACTS BECAME INVERTED DURING TESTING, AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS HAD TO BE PRESSED MULTIPLE TIMES IN ORDER TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTED INTACT AND THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.