FDA Adverse Event Injury Summary report: N

STEM: M-VIZION DISTAL STEM Ø13MM L 140MM STRAIGHT

MDR report key: 19826342 · Received July 25, 2024

Report

Report Number
3005180920-2024-00541
Event Type
Injury
Date Received
July 25, 2024
Date of Event
June 27, 2024
Report Date
July 25, 2024
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030884993
PMA / PMN Number
K170690
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 28 JUNE 2024. LOT 2115443: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 02-FEB-2022. EXPIRATION DATE: 2027-JAN-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL DEVICE INVOLVED: BATCH REVIEW PERFORMED ON 28 JUNE 2024. STEM: M-VIZION 01.22.401 PROXIMAL BODY Ø20MM L 40MM STD WITH HOLES (K201471) LOT 2207732: (B)(4)ITEMS MANUFACTURED AND RELEASED ON 31-OCT-2022. EXPIRATION DATE: 2027-SEP-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CLINICAL EVALUATION PERFORMED BY MEDACTA HIP R&D PROJECT MANAGER: IT IS CONFIRMED THAT THE DISTAL STEM HAS SIGNS OF ABRASION AT THE PROXIMAL PORTION, CAUSED BY A METAL CERCLAGE WIRE THAT WAS DIRECTLY IN CONTACT WITH THE STEM. NO OBVIOUS CAUSES OF STEM-RELATED LOOSENING COULD BE IDENTIFIED, MOST LIKELY THE PREVIOUS FEMUR FRACTURE MAY HAVE COMPROMISED STEM OSSEOINTEGRATION OR THE METAL CERCLAGE WAS NOT STABLE (PREVENTING BONE HEALING). NO ADDITIONAL INVESTIGATION IS NEEDED.

Description of Event or Problem · 0

REVISION SURGERY WAS PERFORMED DUE TO STEM LOOSENING ABOUT 1 YEAR AND 1 MONTH AFTER THE PRIMARY SURGERY. METAL CERCLAGE CAME INTO CONTACT WITH THE STEM CAUSING ABRASION. CERCAGLE WAS USED TO FIX A FEMORAL FRACTURE THAT OCCURRED WITH COMPETITOR COMPONENTS (THE PATIENT HAD A PRIMARY SURGERY WITH COMPETITOR'S COMPONENTS).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2302893 STEM: M-VIZION DISTAL STEM Ø13MM L 140MM STRAIGHT HIP FEMUR DISTAL STEM LZO MEDACTA INTERNATIONAL SA 2115443 07630030884993

Patients

Seq Age Sex Outcome Treatment
1 NA Female Required Intervention