FDA Adverse Event Malfunction Summary report: N

ACCESS® 2 IMMUNOASSAY SYSTEM

MDR report key: 1982590 · Received February 9, 2011

Report

Report Number
2122870-2011-00344
Event Type
Malfunction
Date Received
February 9, 2011
Date of Event
January 11, 2011
Report Date
January 11, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K922823/A007
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER STATED THAT A SERIES OF TEMPERATURE STATUS MONITOR ERROR MESSAGES WERE POSTED TO THE EVENT LOG JUST PRIOR TO THE BURNING SMELL. SERVICE WAS DISPATCHED ON 01/11/2011 FOR THIS EVENT. THE BCI FIELD SERVICE ENGINEER (FSE) DISCOVERED THAT A CHASSIS/COVER SCREW HAD FALLEN ON THE OBSTRUCTION DETECTION PCB BOARD AND SHORTED IT OUT. THE FSE REPLACED THE BOARD AND ASSOCIATED RIBBON CABLE. THE FSE CALIBRATED THE OBSTRUCTION PRESSURE SENSOR AND PERFORMED A ROUTINE SYSTEM CHECK AND QC. ALL RESULTS WERE WITHIN THE SPECIFICATIONS. HARDWARE ISSUE WAS THE ROOT CAUSE FOR THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER, INC. (BCI) IN REGARDS TO A BURNING SMELL COMING FROM ACCESS 2 IMMUNOASSAY SYSTEM. THE CUSTOMER POWERED DOWN THE INSTRUMENT AND REMOVED ALL REAGENTS FOR OFF-BOARD STORAGE. THERE WAS NO FLAME OR FIRE, AND THE FIRE DEPARTMENT WAS NOT DISPATCHED. THERE WAS NO REPORT OF INJURY TO THE USER, AND THE CUSTOMER DID NOT SEEK OR NEED MEDICAL ATTENTION.