FDA Adverse Event Injury Summary report: N

LEVEL ONE

MDR report key: 19825786 · Received July 25, 2024

Report

Report Number
9610905-2024-00042
Event Type
Injury
Date Received
July 25, 2024
Date of Event
July 8, 2024
Report Date
January 16, 2025
Manufacturer
KLS MARTIN SE & CO. KG
Product Code
HTW
UDI-DI
00888118040817
PMA / PMN Number
CLS 1 EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS PERFORMED ON THE BASIS OF COMPLAINT STATISTICS AS NO DEVICE WAS RETURNED FOR EVALUATION. THE COMPLAINT PERCENTAGE WAS CALCULATED, AND IT IS DETERMINED THAT THE COMPLAINT PERCENTAGE FALLS WITHIN THE DESIGN RISK LIMITS ADHERED TO AT KLS SE. DURING THE INVESTIGATION THE PRODUCT LOT NUMBER WAS NOT IDENTIFIED; THEREFORE, THE DEVICE HISTORY RECORDS WERE NOT REVIEWED. THE FAILURE ROOT CAUSE CANNOT BE DETERMINED DUE TO THE DEVICE NOT BEING RETURNED. IF FURTHER INFORMATION IS OBTAINED THAT MIGHT ADD VALUE TO THE CONTENTS OF THE INVESTIGATION REPORT, AN ADDITIONAL FOLLOW-UP REPORT WILL BE SUBMITTED. THE REPORTED RPMS EXCEEDED THE MAXIMUM RECOMMENDED SPEED FOUND IN THE IFU.

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS SUCCESSFULLY COMPLETED. THE RESULTS OF THE INVESTIGATION HAVE IDENTIFIED NO MANUFACTURING ISSUES, NO DESIGN ISSUES AND/OR CORRECTIVE ACTIONS NECESSARY AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED THAT MIGHT ADD VALUE TO THE CONTENTS OF THE INVESTIGATION REPORT, AN ADDITIONAL FOLLOW-UP REPORT WILL BE SUBMITTED. THE REPORTED RPMS EXCEEDED THE MAXIMUM RECOMMENDED SPEED FOUND IN THE IFU. LOTS IDENTIFIED: 33606099 - MFG DATE 04/03/2024; 33606103 - MFG DATE 04/08/2024.

Description of Event or Problem · 0

IT WAS REPORTED DURING INITIAL SURGERY A TWIST DRILL BROKE DURING USE. A PORTION OF IT REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
351082 LEVEL ONE TWIST DRILL HTW KLS MARTIN SE & CO. KG 25-449-16-91 SEE H11 00888118040817

Patients

Seq Age Sex Outcome Treatment
1 27 YR Male Other