FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1982577 · Received January 12, 2011

Report

Report Number
2531779-2011-00185
Event Type
Malfunction
Date Received
January 12, 2011
Date of Event
December 14, 2010
Report Date
December 14, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR INVESTIGATION. EVALUATION CONFIRMED THAT THE PUMP RESET TO THE DEFAULT TIME DURING A BATTERY CHANGE. DURING INVESTIGATION, THE FAILURE WAS CONFIRMED; THE PUMP WAS UNABLE TO HOLD THE DATE/TIME AND HAD RESET ITSELF TO THE DEFAULT TIME DUE TO A LEAKING BT1 (INTERNAL BATTERY).

Description of Event or Problem · 1

THE PATIENT'S MOM INDICATED THAT THE BATTERY WAS REPLACED IN THE PUMP AND THE DATE/TIME SETTING WAS CORRECTED. AT AN UNSPECIFIED TIME AFTER THE DATE/TIME SETTING WAS SET, THE PUMP'S TIME SETTING WAS CHANGED WITHOUT USER INTERVENTION. THERE WAS NO REPORTED ADVERSE EVENT WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 12 YR