FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1982576
·
Received January 12, 2011
Report
- Report Number
- 2531779-2011-00194
- Event Type
- Malfunction
- Date Received
- January 12, 2011
- Report Date
- December 13, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION AND EVALUATED. THE KEYPAD APPEARED TO BE INTACT; HOWEVER, THE OK/CONTRAST KEY CONTACTS WERE INTERMITTENTLY RESPONDING TO USER INPUTS, THE UP/DOWN KEYS REQUIRED EXCESSIVE FORCE TO ACTIVATE DURING TESTING, THE UP/DOWN KEY CONTACTS WERE MISALIGNED, THE UP/DOWN/OK/CONTRAST KEY CONTACTS BECAME INVERTED DURING TESTING, AND THERE WAS EVIDENCE OF CONTAMINATION UNDER THE KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTONS HAD TO BE PRESSED MULTIPLE TIMES IN ORDER TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND THE PUMP HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | IR 1250 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |