FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL TIP 1ML

MDR report key: 19825753 · Received July 25, 2024

Report

Report Number
1911916-2024-00540
Event Type
Malfunction
Date Received
July 25, 2024
Date of Event
July 14, 2024
Report Date
July 30, 2024
Manufacturer
BECTON DICKINSON
Product Code
FMF
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP FOR DEVICE EVALUATION. IT WAS REPORTED THE VOLUME MARKINGS ON SIDE OF SYRINGE ARE MISSING HALFWAY UP THE SYRINGE. TO AID IN THE INVESTIGATION, TWENTY SAMPLES AND ONE PHOTO WAS PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. EIGHTEEN OF THE SAMPLES CAME IN SEALED PACKAGING BLISTERS, ONE SAMPLE CAME IN AN OPENED PACKAGING BLISTER AND ONE SAMPLE CAME WITH NO PACKAGING. A VISUAL INSPECTION WAS PERFORMED, AND THE SYRINGE BARREL SCALE IS PARTIALLY PRINTED. THE PHOTO PROVIDED SHOWS A SYRINGE WITH A PARTIALLY PRINTED SCALE MARKING. NO OTHER DEFECTS OR IMPERFECTIONS WERE OBSERVED. THIS DEFECT COULD OCCUR IF THERE WAS A JAM DURING THE SYRINGE BARREL PRINTING PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309653, LOT 4080174. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. THE SAMPLES WILL BE SHOWN TO ASSOCIATES FOR AWARENESS. BASED ON THE INVESTIGATION AND WITH THE RETURNED ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED.

Description of Event or Problem · 0

MATERIAL: 309653, BATCH#: 4080174. IT WAS REPORTED BY THE CUSTOMER THAT VOLUME MARKINGS ON SIDE OF SYRINGE IS MISSING HALFWAY UP THE SYRINGE. VERBATIM: RCC RECEIVED A COMPLAINT VIA COMMUNITY. VOLUME MARKINGS ON SIDE OF SYRINGE IS MISSING HALFWAY UP THE SYRINGE.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION RECEIVED MATERIAL: 309653, BATCH#: 4080174. IT WAS REPORTED BY THE CUSTOMER THAT VOLUME MARKINGS ON SIDE OF SYRINGE IS MISSING HALFWAY UP THE SYRINGE. VERBATIM: RCC RECEIVED A COMPLAINT VIA COMMUNITY. VOLUME MARKINGS ON SIDE OF SYRINGE IS MISSING HALFWAY UP THE SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
365298 SYRINGE 50ML LL TIP 1ML SYRINGE, PISTON FMF BECTON DICKINSON 4080174

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown