- Report Number
- 1911916-2024-00533
- Event Type
- Malfunction
- Date Received
- July 25, 2024
- Date of Event
- July 11, 2024
- Report Date
- August 1, 2024
- Manufacturer
- BECTON DICKINSON
- Product Code
- GAA
- UDI-DI
- 30382903052005
- PMA / PMN Number
- UNK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PR (B)(4) FOLLOW UP: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 305200 AND LOT NUMBER 2172517. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.
INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
ADDITIONAL INFORMATION RECEIVED WAS THERE ANY PATIENT HARM DUE TO THE ISSUE? THERE WAS NO PATIENT HARM DUE TO THIS ISSUE. MATERIAL #:305200; BATCH#:2172517. IT WAS REPORTED BY CUSTOMER THAT THE NEEDLE SHOT OFF SYRINGE AND DRUG WENT BLASTING OUT, WAS NOT ABLE TO INJECT. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. NEEDLE SHOT OFF SYRINGE AND DRUG WENT BLASTING OUT, WAS NOT ABLE TO INJECT. PRODUCT#: 305200; LOT#: 2172517.
MATERIAL #:305200 BATCH#:2172517 IT WAS REPORTED BY CUSTOMER THAT THE NEEDLE SHOT OFF SYRINGE AND DRUG WENT BLASTING OUT, WAS NOT ABLE TO INJECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2303796 | NEEDLE FILTER 19X1-1/2 TW | NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE | GAA | BECTON DICKINSON | 2172517 | 30382903052005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |