FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 1982398 · Received January 14, 2011

Report

Report Number
9615739-2011-00007
Event Type
Malfunction
Date Received
January 14, 2011
Date of Event
December 20, 2010
Report Date
December 20, 2010
Manufacturer
HILL-ROM INDUSTRIES S.A.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE CASTERS AT THE FOOT END OF THE BED NOT HOLDING AND WERE RATCHETING. HE ATTEMPTED TO ADJUST THE CASTERS FIRST, BUT FOUND THE BRAKE ON THE CASTER WAS WORN. THE CUSTOMER IS CONSTANTLY PLACING THE BRAKE ON/OFF AND MOVING THE BED FROM THE FOOT END TO PLACE THE PATIENT IN AND OUT OF BED. THE TECHNICIAN REPLACED THE CASTERS ON THE FOOT END OF THE BED TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED WHEN THE BRAKES ARE SET ON THE BED, THE BRAKES ARE NOT HOLDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAREASSIST BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INDUSTRIES S.A. 1170

Patients

Seq Age Sex Outcome Treatment
1 UNK