FDA Adverse Event Malfunction Summary report: N

POWERED WHEELCHAIR

MDR report key: 1982136 · Received January 14, 2011

Report

Report Number
1525712-2011-00014
Event Type
Malfunction
Date Received
January 14, 2011
Date of Event
August 17, 2010
Report Date
January 13, 2011
Manufacturer
INVACARE
Product Code
ITI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEALER WAS SERVICING THEIR CHAIR WHEN THEY HEARD A POP AND OBSERVED A PUFF OF SMOKE EMIT FROM THE JOYSTICK. THE JOYSTICK WAS RECEIVED AND INSPECTED. NO EXTERNAL DAMAGE WAS OBSERVED. THE JOYSTICK WAS CONNECTED TO A WHEELCHAIR POWERED ON AND OPERATED WITH NO DISCREPANCIES. INTERNAL INSPECTION OF THE JOYSTICK REVEALED A DAMAGED CAPACITOR. ALL DAMAGE TO THE COMPONENT WAS CONTAINED WITHIN THE METAL ENCLOSURE OF THE JOYSTICK. NO RISK OF SHOCK IS PRESENTED TO THE USER. AS IN ANY ELECTRONIC COMPONENT FAILURE, SOME DEGREE OF SMOKING CAN OCCUR; HOWEVER, THIS IS NOT AN INDICATION OF SUSTAINED COMBUSTION. THE DEVICE HAS BEEN REPLACED AND THE CHAIR REMAINS IN USE.

Description of Event or Problem · 1

AFTER INSTALLING A NEW JOYSTICK ON THE CHAIR, THE DEALER ALLEGEDLY SAW A PUFF OF SMOKE COME OUT OF THE CHARGER PORT. NO INJURY IS ALLEGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POWERED WHEELCHAIR 890.3860 ITI INVACARE MK6-MPJC

Patients

Seq Age Sex Outcome Treatment
1