INFANT CONTINUOUS FLOW BREATHING CIRCUIT
Report
- Report Number
- 9611451-2011-00061
- Event Type
- Malfunction
- Date Received
- February 8, 2011
- Report Date
- January 11, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K020332
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE IS NOT EXPECTED TO BE RETURNED TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. OUR ANALYSIS IS ACCORDINGLY BASED ON THE DESCRIPTION OF EVENTS, ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER AND OUR KNOWLEDGE OF THE PRODUCTS. A LOT CHECK REVEALED NO OTHER COMPLAINT OF THIS NATURE FOR THIS LOT NUMBER. THE RT329 INFANT CONTINUOUS FLOW BREATHING CIRCUIT KIT INCLUDES A PRESSURE MANIFOLD WHICH ENSURES PATIENT SAFETY BY LIMITING THE PRESSURE DELIVERED IN AN EVENT OF AN OCCLUSION, AS WELL AS ALLOWING CONNECTION TO A PRESSURE MONITORING DEVICE OR AN OXYGEN ANALYZER. THE MANIFOLD IS PACKED WITH THE BLUE CAP IN PLACE. THE PRESSURE MANIFOLD IS PRESSURE TESTED AND VISUALLY INSPECTED PRIOR TO LEAVING OUR FACILITY. ANY OF THOSE THAT FAIL THE PRESSURE TEST ARE DISCARDED. THEREFORE THIS SUGGESTS THAT IT IS LIKELY THAT THE CAP BECAME LOOSE POSTPRODUCTION DURING TRANSPORT. OUR USER INSTRUCTIONS STATE THE FOLLOWING: "CHECK THAT ALL CONNECTIONS, CAPS AND/OR PLUGS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."
A HOSPITAL IN (B)(6) REPORTED THAT THE BLUE PLUG ON THE PRESSURE MANIFOLD OF AN RT329 INFANT CONTINUOUS FLOW BREATHING CIRCUIT KIT "EITHER CAME OUT OR WAS LOOSE THEN CAME OUT OVER TIME" FROM THE CHAMBER PORT AFTER ONE DAY OF USE. THE HOSPITAL REPORTED THAT THERE WERE NO PHYSICAL DAMAGE TO THE CHAMBER OR PRESSURE MANIFOLD. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT CONTINUOUS FLOW BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LTD | RT329 | 101020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |