CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2011-00203
- Event Type
- Death
- Date Received
- February 8, 2011
- Date of Event
- October 15, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- RISK MANAGER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) NO ANOMALIES FOUND, PROXIMAL CONDUCTOR STRETCHED, BLOOD/BODY FLUID (NOT OBSTRUCTED), INNER INSULATION TORN, OUTER INSULATION COSMETIC AND BREACHED CUT, LEAD STRETCHED, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4) NO ANOMALIES FOUND.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY (B)(4) NO ANOMALIES FOUND, PROXIMAL CONDUCTOR STRETCHED, BLOOD/BODY FLUID (NOT OBSTRUCTED), INNER INSULATION TORN, OUTER INSULATION COSMETIC AND BREACHED CUT, LEAD STRETCHED, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED.
IT WAS NOTED THE PATIENT DIED APPROXIMATELY 7 MONTHS FOLLOWING DEVICE IMPLANT. THE CAUSE OF DEATH IS BEING REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR | Death | LEAD/STJ IMPLANTABLE PACING LEAD| LEAD/STJ IMPLANTABLE PACING LEAD |