AUTOSOFT XC
Report
- Report Number
- 3003442380-2024-16154
- Event Type
- Malfunction
- Date Received
- July 24, 2024
- Date of Event
- June 15, 2024
- Report Date
- November 6, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244016651
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS H11: INVESTIGATION SUMMARY COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6003992, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6003992 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 108 AND MANUFACTURED IN THE LINE L5 ON 07-NOV-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 3L00038 WAS MANUFACTURED ACCORDING TO THE WI VERSION 10 AND MANUFACTURED IN THE MACHINE IGTL01, ON 03-NOV-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY: GLUING OF TUBING OF THE LOT 3L00039 WAS MANUFACTURED ACCORDING TO THE WI VERSION 10 AND MANUFACTURED IN THE MACHINE IGTL01, ON 05-NOV-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
UNOMEDICAL REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES IT WAS REPORTED THAT THE PATIENT FACED A BLOCKAGE EVENT WITH AN INFUSION SET WHERE THE TUBING WAS BLOCKED AND WAS ALERTED BY ALARM 2 FOLLOWED BY ALARM 26. THIS LED TO INCREASE IN THE BLOOD GLUCOSE LEVEL AND THE PATIENT TOOK THE CORRECTION INJECTION VIA MULTIPLE DAILY INJECTIONS TO MANAGE THE SAME. ALSO, PATIENT CHANGED SUPPLIES AS NECESSARY AND RESUMED INSULIN THERAPY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 102997 | AUTOSOFT XC | UNO INSET I 60/9 GREY TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1001681 | 6003992 | 05705244016651 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Male |